1. Equipment introduction
Water flow standard device is a standard equipment for verifying various types of water flow meters and researching flow meter test methods.
2. According to the standard
JJG 164-2000 Standard Facilities for Liquid Flowrate
JJG 198-1994 Verification Regulation of Velocity Flow Meter
JJG 640-1994 Verification Regulation of differential pressure type flowmotor
JJG 643-2003 Flow Standard Facilities by Master Meter Method
JJG 667-1997 Liquid Positive Displacement Flow Meter
JJG 1029-2007 Vortex-shedding Flowmeter
JJG 1033-2007 Electromagnetic Flowmeters
JJG 1038-2008 Goriolis Mass Flow Meters
3. Technical parameters
Medium: Water
Pipeline internal diameter and flow: Reference vortex flowmeter flow range
Verification caliber range: DN15-DN300
Static mass method design flow range: 0.5-720 m3/h
Master meter method design flow range: 0.5-1800m3/h
Flowmeter to be calibrated: Coriolis mass flowmeter, turbine flowmeter, vortex flowmeter, mass flowmeter, electromagnetic flowmeter
Methods: Static mass method, dynamic master meter method. The mass method and master meter method are combined.
Static mass method uncertainty: 0.05%
Master meter method uncertainty: 0.3%
System pressure fluctuations: ≤0.3%
Flow stability: 0.2%
Automatic data acquisition, data processing, printing; Automatic valve control and automatic flow adjustment
The device is tiled. Set 4 clamp meter epitopes : DN50 epitope is compatible with DN40, DN25, DN15; DN100 epitope is compatible with DN80 and DN65, DN200 epitope is compatible with DN150 and DN125; DN300 epitope is compatible with DN250. Moving bracket is used for the replacement of test pipeline, which is convenient and flexible.